Search 
Troubleshooting
Questionnaire
Precautionary Measures while running the ELISA test
  1. Allow the plate to equilibrate at room temperature for 5 to 10 minutes prior to starting the assay.
  2. Carry out the assays at approximately 25oC.
  3. Ensure that wash apparatus is working correctly, recalibrate if necessary.
  4. Change pipette tips between each standard, sample or reagent. Use separate reservoirs each reagent.
  5. Wells should appear dry after aspiration.
  6. Ensure adequate mixing of reagents.
  7. Adhere to recommended incubation periods and equipment as far as possible.
  8. Once started, the assay should be completed. Have all standards / samples prepared appropriately before commencement of the assay.
  9. Prepare substrate(s) within the time recommended in the kit insert.
  10. Ensure that all reagents are at room temperature before pipetting into the wells unless otherwise instructed in the kit insert.
Inadequate color development
  1. Possibility of inadequate volume of substrate added to the wells.
  2. Incomplete incubation of plate after adding the substrate solution.
  3. The conjugate concentration may be different than that recommended.
No color development
  1. Errors in incubation times or temperature changes may lead to absence of color development.
  2. It is necessary to read the plate at the correct calorimetric wavelength within 15 minutes after adding the stop solution.
High Background
  1. Incomplete washing of wells with less than the recommended amount of buffer can result in high background. Please follow the washing instruction given in the insert.
  2. Incubation of plate for very long time or higher than recommended temperature can result in high background.
Drift
  • It is possible due to interrupted assay set up or reagents not at appropriate room temperature.
Poor precision
  1. Unequal washing of wells during washing steps.
  2. Inadequate mixing of reagents.
  3. Wells should be aspirated carefully. Tap the plates wells down on paper towels after the last wash to ensure residual moisture is removed.
  4. Do not prepare the substrate too early. Follow the time recommended in the kit insert.
Poor standard curve (Quantitative assay)
  1. While loading the antibodies in the plate, avoid creating bubbles in the wells.
  2. If the % CV values for the standard curve OD values are consistently above 15% pay particular attention to pipetting technique. If standard curve ODs are acceptable but sample precision is not, follow the recommendations to improve precision.
  3. Washing the plate between incubation steps should be uniform and each well should be filled completely with wash buffer.
Edge effect
  1. Fluctuations in temperature and humidity around the work surface can lead to edge effects.
  2. Edge effects could also arise due to inadequate fixing of plate covers, leading to evaporation.
© 2006, All rights reserved